Click to Return to Home Page
Log In | Register Now

       New to Our Site?

















 


   

UV Calibration Kit

UV-vis Calibration Kit™ Solutions and Concentrations

Absorbance Solution No. 1. nominal 0.02 mg/mL Potassium Dichromate
30ml bottle
Absorbance Solution No. 2. nominal 0.04 mg/mL Potassium Dichromate
30ml bottle
Absorbance Solution No. 3. nominal 0.06 mg/mL Potassium Dichromate
30ml bottle
Absorbance Solution No. 4. nominal 0.08 mg/mL Potassium Dichromate
30ml bottle
Absorbance Solution No. 5. nominal 0.10 mg/mL Potassium Dichromate
30ml bottle
Absorbance Solution No. 6. nominal 0.12 mg/mL Potassium Dichromate
30ml bottle
Absorbance Solution No. 7. nominal 0.14 mg/mL Potassium Dichromate
30ml bottle
Reference Solution Diluent 0.001N perchloric Acid used to prepare linearity standards
60ml bottle
Wavelength Calibration Solution*, 4% Holmium Oxide
30ml bottle
Stray Light Solution, 1% Potassium Iodide in water
30ml bottle

Also included with the kit is an Excel Template for ease of qualification, data recording and reporting of the actual data and acceptance criteria.

* The Holmium Oxide solution is chemically identical to NIST SRM 2034 and is directly qualified against NIST standards during manufacture.

See Next Page for Principles of Operation

 