HPLC Application Sheet
CARVONE - Cogent UDC-Cholesterol™ Retention Map
4 Different Mobile Phases



Method Conditions



Column Cogent UDC™ Cholesterol
Catalog No. 69069-15R
Dimensions 150mm x 4.6mm ID
Mobile Phase Water/THF
Water/Acetonitrile
Water/Methanol
Hexane/Isopropanol
Flow Rate 1mL/minute
Sample Carvone
Synonym 2-cyclohexen-1-one, 2-methyl-5-(1-methylethenyl)methyl-5-(1-methylethenyl)
Injection Volume 10µL
Detection 255nm

Click Here to view stationary phase specifications
Click Here to view chemical structures of analytes

Discussion of Chromatogram:
The retention characteristics of Carvone on this UDC-Cholesterol Column in aqueous reverse phase mode is typical of a C18 or C8 column. The retention times change respectively with THF being a stronger solvent than Acetonitrile which is stronger than Methanol (shown).

This column can also be used in either Aqueous-normal phase or Organic-normal phase (shown above). Other phases such as Cyano or Amino are not optimal choices if you want to switch between Aqueous-reverse phase and Organic normal phase. UDC-Cholesterol and other Cogent TYPE-C HPLC Columns are unique as they require relatively minimal re-equilibration effort to switch between Aqueous-normal phase and Reverse phase and back to Organic-normal phase.