![]() |
![]() |
Log In | Register Now New to Our Site? |
||||||||||||||||||||||
|
||||||||||||||||||||||||
|
A Generic Method for Scouting with Cogent TYPE-C Silica Based HPLC Columns
It is easy to use a Cogent HPLC column made with TYPE-C Silica. The power of these columns is found in the wide range of solvents they can be used with and the unique and strong retention they can have for polar and non polar compounds. Selecting which TYPE-C column to use (C18, C8, Cholesterol or Silica-C) will be determined by your compounds of interest. Mixtures that will be highly polar and do not contain non polar compounds might be better suited by the Silica-C where compounds with both polar and non polar might be better suited to the Bidentate C8, C18 or UDC Cholesterol. Step 1. After you have properly installed and conditioned the column it is a good idea to start with a typical reverse phase gradient run. We suggest starting with an acidified mobile phase of Water as component A and Acetonitrile as component B. Acidify both components with up to 0.5% of an acid such as Formic, Methyl Phosphonic or Acetic Acid. If you are not using LCMS, TFA (up to 0.1%) is another good candidate. Step 2. Run about 6 column volumes of the mobile phase in Step 1 at 95% Water. Step 3. Set up your instrument to run a shallow gradient from 95% Water (A) to 40% Water (A) over 20 minutes on a 75mm long column. For longer or shorter columns, increase the gradient time proportionally. This long and shallow gradient will be very beneficial for determining the optimal gradient or isocratic method to run your mixture with later. For sharper peaks and less retention, run a shorter (Steeper) gradient from the same starting points to end points. Step 4. Equilibrate the column by running 100% Acetonitrile for approximately 2 minutes for the 75mm long column. Step 5. Set up your instrument to run a shallow gradient using the same mobile phase to run from 90% Acetonitrile (B) to 40% Acetonitrile(B) over 20 minutes for a 75mm long column. For longer columns increase the gradient time proportionally. This long and shallow gradient will be very beneficial for determining the optimal gradient or isocratic method to run your mixture with later. For sharper peaks and less retention, run a shorter (Steeper) gradient from the same starting points to end points. Step 6. Evaluate both gradient runs for retention time, peak shape and elution order. Since analyte retention on these columns is compound and method specific some compounds may not retain in Step 3 (Reverse Phase) and some may not retain in Step 5 (Aqueous Normal Phase). However, one column could produce an isocratic run which retains both polar and non polar compounds. Note: The Cogent Bidentate C8, C18 and UDC-Cholesterol columns have a unique quality in that they may retain polar compounds not retained on other columns while run at 100% Water without loss of retention with continued use. You could insert an isocratic run at 100% acidified Water after Step 3 and before Step 4. Optimizing Suggestions: If you do not find satisfactory results or once you have established some retention and selectivity, you may need to optimize the method for your specific compounds. We have listed some suggestions below that will offer some ideas. If you are not satisfied with the results, please contact our technical support team.
|
|||
| [TOP] |
|
Privacy Statement Copyright © 2005, MicroSolv Technology Corporation. Site developed by Net Ink Designs. |