Column Size and Flow Rates
|
FLOW RATES FOR USE WITH JORDI GPC/SEC COLUMNS
The 10mm x 500mm Jordi GPC columns is larger than the typical 7.8mm x 300mm GPC Columns.
- Larger Injection Volumes are possible with Jordi Columns without overload.
- Column flow rates for Jordi columns are typically 1.5mL/minute. Higher plate counts can be obtained at
lower linear velocities.
- Normal analysis times are about 22 minutes at 1.5mL/minute which can be cut into half when run on a Jordi
column at 3.0mL/minute with minimum loss of plate count.
For methods run at 0.5mL/minute, the plate counts are significantly higher especially for the more narrow dispersed
higher polymers.
The most optimal flow rate is always the lowest possible when analyzing polymers. This
may not be true for small molecules.
If you change flow rate to increase efficiency for a narrow standard, a more narrow peak will result. If the
standard is dispersed, it will either exhibit a broader peak as the different MW species start to separate, or it
will remain somewhat the same.
When separating a polymer distribution, it is generally believed that the optimum efficiency can be determined by
measuring the plate count of an unretained molecule such as toluene at different flow rates. This will not be true
for polymer separation as the optimal flow rate is much lower than that of a smaller molecule.
See the chart above: Higher resistance to mass transfer of the polymer as illustrated above by the Van Deemter
Plots.