Click to Return to Home Page
Log In | Register Now

       Help



















 


 


HPLC Application Sheet
Steriods Mixture (3) - Cogent UDC-Cholesterol™

Chromatogram A
Chromatogram B
Chromatogram C

PDF
Click here to view
printable Application Sheet

Method Conditions



Mobile Phase   
  Chromatogram A:  60% Aqueous (0.1%TFA)
40% Acetonitrile (ACN)
Chromatogram B:  45% Aqueous (0.1%TFA)
55% MeOH
Chromatogram C:  56% Aqueous (0.1%TFA)
24% MeOH and 20% ACN
Column Cogent UDC-Cholesterol™
Catalog No. 69069-75R
Dimensions 75 x 4.6mm id
Flow Rate 1.0ml/min
Detection UV, 240nm
Temperature 15°C


Discussion of Chromatograms


An FDA requirement for Birth Control Product Analysis is the resolution of hormonal steroids ETES and ESTN.
  • With Cogent Bidentate C18™ and Cogent UDC-Cholesterol™ HPLC columns, peaks ETES and ESTN are resolved with polarity-based, Reverse Phase, ACN as the organic modifier.
  • With other C18 columns, these same peaks are not resolved.
  • When eluting with the shape recognition in Methanol on the Cogent UDC Cholesterol™ column, ETES and ESTN are extremely well resolved. However, in this test mix ETES now co-elutes with ESTD.
  • By mixing the selectivity mechanisms, polarity (ACN) and shape recognition (MeOH), partial resolution was achieved on all compounds using the UDC Cholesterol™ column. Scaling up to a 250x4.6mm column resulted in baseline resolution of all peaks.

Implications, Advantages & Uses
Using a single UDC Cholesterol™ column with orthogonal selectivity, polarity in ACN and shape recognition in Methanol can lead to unique problem solving.





[TOP]


Privacy Statement             Copyright © 2005, MicroSolv Technology Corporation.             Site developed by Net Ink Designs.