Click to Return to Home Page
Log In | Register Now

       Help



















 


 


Reproducibility, Run to Run
Reproducible retention of aromatic amino acids using a Cogent Silica-C™ column and a simple LC-MS compatible mobile phase.

charts
This is six chromatograms
laid on top of each other.


PDF
Click here to view
printable Application Sheet
Note: The mixture of amino acids was prepared and injected without any derivatization. The mobile phase used is LC-MS compatible.


Method Conditions


Column Cogent Silica-C™, 4µm, 100A
Catalog No. 40000-75P
Dimensions 4.6 x 75 mm
Solvents A: DI water + 0.1 % formic acid
B: acetonitrile + 0.1% formic acid
Mobile Phase 80%B/20%A
Flow rate 1.0 mL/minute (t0 = 0.85 min)
Peaks 1. L-(+)-alpha-phenylglycine
2. L-phenylalanine
Injection Volume  2 µL
Sample Matrix 0.3 mg/mL of each sample dissolved in 50% acetonitrile/50% DI water + 0.5% formic acid.
Detection 254 nm UV


Discussion


Two important aromatic amino acids: L-(+) – alpha - phenylglycine and L – phenylalanine were analyzed using a Cogent Silica-C™ column and an ANP (aqueous normal phase) mobile phase. The retention of the amino acids was remarkably reproducible for more than 20 injections. For clarity purposes on the chromatogram presented only 6 consecutive injections are overlaid. The calculated %RSD for 20 injections were around 0.3%. Conclusion: these columns are not only very reproducible but extremely stable.






[TOP]


Privacy Statement             Copyright © 2005, MicroSolv Technology Corporation.             Site developed by Net Ink Designs.