HPLC Application Sheet
Polycyclic Aromatic Hydrocarbons (PAH): A Reverse Phase Separation

charts

Note: Cogent Bidentate C18 with Selectivty3™ operates in three modes of chromatography (reversed phase, aqueous normal phase and organic normal phase). Separation of PAHs illustrates its excellent capabilities in the reversed phase mode.

PDF
Click here to view
printable Application Sheet



Method Conditions



Column Cogent Bidentate C18, 4µm, 100A
Catalog No. 40018-75P
Dimensions 4.6 x 75mm
Mobile Phase Acetonitrile/DI Water 70:30
Flow rate 0.5 mL/minute
Injection Volume  1 µL
Samples
  • Naphthalene
  • Phenanthrene
  • Fluorene
  • Pyrene
  • Impurities or decomposition products - 1mg of
    each sample was dissolved in 1 mL of the mobile
    phase
  • Detection UV at 254 nm


    Discussion


    PAH (Polycyclic Aromatic Hydrocarbons) determination in several matrices (soil, food, air, body fluids etc.) is a topic of interest for routine quality control or screening analyses. Cogent Bidentate C18™ columns offer good resolution, reproducible retention time and peak shape for PAH compounds. In addition to the main components, small peaks due to impurities or decomposition products are well resolved.