HPLC Application Sheet
Bidentate C8 column for more Hydrophobic compounds
Shorter Analysis Time
Method Conditions
Discussion For separations requiring a less hydrophobic stationary phase Cogent C8 can be used successfully as demonstrated in the chromatograms. The retention on the C8 column is lower when compared with a C18 column under the same mobile phase conditions, but since the efficiency is excellent the separation goals can be achieved in a shorter time. The C8 phase is suited for more hydrophobic proteins and peptides and large nonpolar organic compounds. For increased retention higher percentages of water in the mobile phase can be used. The diminished hydrophobicity is shown in Table 1. | ||||||||||||||||||||||||||||||||||||||