Precise Determination of Atropine using a simple gradient
      No Ion Pair Reagent Necessary

charts
PDF
Click here to view
printable Application Sheet

Note: After a simple sample clean up procedure, the method can be applied for monitoring atropine concentrations in biological specimens in cases of drug poisoning. The recoveries of atropine added to drug-free specimens which were analyzed using the described method were satisfactory with coefficients of variation of 4% or less.
Method Conditions



Column Cogent Bidentate C18™, 4µm, 100A
Catalog No. 40018-75P
Dimensions 4.6 x 75 mm
Mobile Phase A: DI water + 0.1 acetic acid + 0.005% TFA
B: Acetonitrile + 0.1% acetic acid + 0.005% TFA
Both solutions were vacuum filtered through a 0.45µm nylon filter
Flow Rate 1 µL/min.
Gradient 0 min - 10% B,
0 to 4 min to 30%B
4 – 6 min hold 30%B
6.01 min back to 10%B
Injection 1 microL
Sample Prepared in 50% solution A/50% solution B, concentration 1 mg/mL and was filtered through a 0.45 µm nylon membrane
Peak Atropine
Injection 1: RT = 4.772 min, Injection 2: RT = 4.773 min,
Injection 3: RT = 4.772 min Injection 4: RT = 4.774 min
Detection UV 214 nm

Discussion

Chromatographic separation and quantification methods of tropa alkaloids are often described in the literature and the method of choice is usually ion-pair chromatography (IPC), which requires long equilibration times and it is not very robust. This method shows a symmetrical peak for atropine using a simple gradient method that does not include any ion pair reagents which can cause damage to columns and lack eproducibility. The retention times are extremely repeatable but one of the best advantages to this method is the time savings between runs. Using a Cogent Bidentate C18 column the equilibration time was very short (5 min between every gradient run).