Controlled EOF and Zero EOF Coated/Bonded Capillaries for CE & cIEF
These 1-meter capillaries are coated with various amounts of our proprietary polymer, which exposes Sulfonic acid
groups to the lumen of the capillary in the controlled EOF Capillary. The varying amounts of polymers provides
controlled amount of resulting EOF. They are pH independent in that the EOF does not change when you change the pH
of the run buffer. The Zero EOF capillaries are neutral coated (bonded) with linear polyacrylamide (LPA) which does
not produce an EOF. They are durable and should give 80 hours of life if used according to the products specifications.
Specifications:
The pH range of the capillaries is 2-8.5
Column ID: 50m or 75m
Column OD: 365µm
Capillary Exterior Coating: Polyimide
Length: 1 meter
|
Electro Osmotic Flow: pH 7.5
Bare Silica
High Flow
Medium Flow
Low Flow
Zero Flow
|
7.0 x 10-4 cm2/volt second
5.5 x 10-4 cm2/volt second
3.5 x 10-4 cm2/volt second
2.5 x 10-4 cm2/volt second
No EOF Detected
|
|
Relative Migration Times of A Neutral Marker
Controlled v. Bare Fused Silica Capillary
Bare Fused Silica
High
Med
Low
Zero
|
3.2 minutes
4.3 minutes
6.5 minutes
8.2 minutes
undetected
|
|
Column Conditioning Procedure:
Never Use Strong Bases on this Capillary
- Carefully remove the capillary from the shipping/storage container. It is fragile.
- Remove protective polyimide coating by using 98% fuming sulfuric acid. One drop should be sufficient for
complete removal. A low temperature flame may be used as well. Always use appropriate and approved safety
precautions when performing the above.
- Condition the capillary by running your “run buffer” through the capillary for 30 minutes.
- Rinse the column (after step 3) with CEwaterTM for 5 minutes.
- Final rinse should be with your run buffer for an additional 10 minutes.
- Repeat this procedure between each injection with reduced run times.
Column Storage Procedure:
- Rinse the column for 25 minutes with your run buffer.
- Rinse the column with CEwaterTM for fifteen minutes.
- If storing for an extended period, we recommend that N2 or He should be blown, under slight pressure, through
the column.