  CElixir Plus - Cations + Amines Kit
Click here to return to
the CElixir home page

 

 
Chart
 
PDF
Click here to view
printable Application Sheet

Separation of Alkali and Alkaline Earth Metals Test Mixture

CElixir-plus™ kit for cations and aliphatic amines is a valuable tool and preferable to traditional methods to separate these compounds such as Ion Chromatography. Easy to use and reliable, save time when buffers are pre-made and quality control tested.

Method Conditions:
Voltage:
30kV
Capillary: MicroSolvCE™ Bare Fused Silica, 75µm x 60cm
Injection: 5 sec @0.5psi
Run Buffer: CElixir-plus™ kit
Detection: Indirect UV 200nm
Polarity: Normal
Instrument: Beckman® P/ACE MDQ™

Method:
Using a Beckman P/ACE MDQ and a 75µm bare fused silica capillary, the earth metals are easily and reproducibly separated using “inverse detection” mode. Buffers are pre-made and supplied in a kit as CElixir-plus™. To avoid carry-over of Sodium, the capillary should be rinsed between runs with a Lithium solution. If Lithium is an analyte of interest, NaOH solution should be used instead.

Peaks:
NH4, Na, K, Li, Mg, Ca (approximately 10mg/L)

Discussion and Rationale on using Inverse Detection:
CE instruments are usually equipped with a UV detector, the problem is that the analyzed cations do not absorb sufficient amount of the UV energy to allow photometric detection. For this reason, an indirect detection technique is used where the supplied buffers (CElixir-plus™) absorbs the UV and the analytes appear as negative peaks. The software of the CE instruments will invert the signal to obtain a normal looking e-gram.